Old 01-11-2017, 12:00 PM   #1
Junior Member
 
Join Date: Jan 2017
Location: Texas
Posts: 8
Rewards Points: 16
Default Klein CL2300 Clamp, Can't Zero DC Amps

I purchased a Klein CL2300 meter. Instructions show how to “zero” the DC Amps. But, it won’t zero. It shows -16 amps. Seems to work fine when I use it on welders leads. I’ve done this in the past with analog clamp gauges. Now, with digital, everything pretty well works the same, but can’t zero. I don’t remember having to zero the analog types. I’m not a professional electrician, I’m a welding inspector. I’ve done this a million times in the past, but “zeroing” a digital meter is all new to me.
1. Should it show “zero”, when I zero it?
2. Is it normal that it “not” show zero? Is it picking up stray signals?
3. If this is normal, do I have to “add” that -16 amps to my final positive readings?
4. What should I have for lunch?
Bob
SlingBob is offline   Reply With Quote
Join Contractor Talk

Join the #1 Electrician Forum Today - It's Totally Free!

ElectricianTalk.com - Are you a Professional Electrical Contractor? If so we invite you to join our community and see what it has to offer. Our site is specifically designed for you and it's the leading place for electricians to meet online. No homeowners asking DIY questions. Just fellow tradesmen who enjoy talking about their business, their trade, and anything else that comes up. No matter what your specialty is you'll find that ElectricianTalk.com is a great community to join. Best of all it's totally free!

Join ElectricianTalk.com - Click Here JOIN FOR FREE


Warning: The topics covered on this site include activities in which there exists the potential for serious injury or death. ElectricianTalk.com DOES NOT guarantee the accuracy or completeness of any information contained on this site. Always use proper safety precaution and reference reliable outside sources before attempting any construction or remodeling task!

Sponsored Links
Advertisement
 
Old 01-11-2017, 12:00 PM   #2
Junior Member
 
Join Date: Jan 2017
Location: Texas
Posts: 8
Rewards Points: 16
Default

Oh yeah... I did read the instructions,,,, and followed them....
SlingBob is offline   Reply With Quote
Old 01-11-2017, 12:22 PM   #3
:-)
 
daveEM's Avatar
 
Join Date: Nov 2012
Location: Alberta
Posts: 2,943
Rewards Points: 940
Default

Klein makes really good pliers... or they used to. I think the pliers arestill pretty goog actually.
99cents likes this.
__________________
<<
They say if opportunity isn't knocking,
...Build a door.
I like perfect.
daveEM is online now   Reply With Quote
Old 01-11-2017, 12:33 PM   #4
Junior Member
 
Join Date: Jan 2017
Location: Texas
Posts: 8
Rewards Points: 16
Default

That's the reason I went with Klein, they always made good pliers.... ;-D
For the limited use I have for a clamp meter, I went a little cheaper than Fluke...
SlingBob is offline   Reply With Quote
Old 01-11-2017, 12:44 PM   #5
Senior Member
 
Join Date: Dec 2016
Location: Virginia
Posts: 302
Rewards Points: 24
Default

Quote:
Originally Posted by SlingBob View Post
That's the reason I went with Klein, they always made good pliers.... ;-D
For the limited use I have for a clamp meter, I went a little cheaper than Fluke...
You get the quality you pay for. We buy our guys Fluke and never have issues unless they are damaged during shipping
MCasey is offline   Reply With Quote
Old 01-11-2017, 01:04 PM   #6
Senior Member
 
John Valdes's Avatar
 
Join Date: May 2007
Location: Pleasure Dome
Posts: 12,674
Rewards Points: 1,240
Default

Do you mean it shows minus 16 amps? You typed -16.
I would put it on a known current carrying conductor and see if its accurate.
I used the DC function on my clamp many times. Never saw zero with clamp empty.
Close but not zero.
Thats the price you pay for digital technology.
John Valdes is offline   Reply With Quote
Old 01-11-2017, 01:09 PM   #7
Petulant Amateur
 
99cents's Avatar
 
Join Date: Aug 2012
Location: Perky Nipples, Canada
Posts: 14,873
Rewards Points: 9,938
Default

Quote:
Originally Posted by SlingBob View Post
Oh yeah... I did read the instructions,,,, and followed them....
You're lucky. I can't read Chinese.
splatz likes this.
99cents is online now   Reply With Quote
The Following 2 Users Say Thank You to 99cents For This Useful Post:
MCasey (01-11-2017)
Old 01-11-2017, 01:10 PM   #8
Junior Member
 
Join Date: Jan 2017
Location: Texas
Posts: 8
Rewards Points: 16
Default

Yes, it is -16 amps (negative). I'm not sure if it is supposed to show dead on zero as does a volt meter, or whether it picks up stray signals when zero'd and empty. Or, maybe it's my magnetic personality. I've messed with this for a couple of weeks, and searched the internet to no avail. Whether I'm in my pickup, walking around in a field, or setting in my recliner, it doesn't go to zero when I press the zero button. It does beep, but I need to know I've got accuracy, and not just a cute R2D2 (star wars?) beep.
SlingBob is offline   Reply With Quote
Old 01-11-2017, 03:02 PM   #9
Senior Member
 
Join Date: Apr 2016
Location: Northern CA
Posts: 313
Rewards Points: 34
Default

When you reverse the direction of the clamp on a wire carrying a known current do you get the same reading (but negative) or one differing by 32A?

Sent from my XT1585 using Tapatalk
Frank Mc likes this.

Last edited by inetdog; 01-11-2017 at 03:29 PM.
inetdog is offline   Reply With Quote
Old 01-11-2017, 03:18 PM   #10
Junior Member
 
Join Date: Jan 2017
Location: Texas
Posts: 8
Rewards Points: 16
Default

Quote:
When you reverse the direction of the clamp on a wire carrying a known current do you get the same reading or one differing by 32A?
Good Question, I haven't paid any attention to that.... That'll be my next trial.... Why didn't I think of that...? I will advise when I find out...
And thank you...
SlingBob is offline   Reply With Quote
Old 01-11-2017, 04:10 PM   #11
Junior Member
 
Join Date: Jan 2017
Location: Texas
Posts: 8
Rewards Points: 16
Default

Well, as a matter of fact.... There does appear to be about 30 amps different when I turn it around on the same cable.... phunky crap.... I think I'm gonna shop for a Fluke...
SlingBob is offline   Reply With Quote
The Following User Says Thank You to SlingBob For This Useful Post:
Flyingsod (01-11-2017)
Old 01-11-2017, 06:09 PM   #12
Senior Member
 
Join Date: Apr 2016
Location: Northern CA
Posts: 313
Rewards Points: 34
Default

Quote:
Originally Posted by SlingBob View Post
Well, as a matter of fact.... There does appear to be about 30 amps different when I turn it around on the same cable.... phunky crap.... I think I'm gonna shop for a Fluke...
That finding proves beyond any doubt that there is a problem with zeroing.
Whether that is a defect in one unit or a problem with the design will probably remain unknown unless you make a warranty claim.

Sent from my XT1585 using Tapatalk
inetdog is offline   Reply With Quote
Old 01-12-2017, 11:33 AM   #13
Junior Member
 
Join Date: Jan 2017
Location: Texas
Posts: 8
Rewards Points: 16
Default Problem solved (maye temporarily)

Klein suggested taking it to any vendor that deals in the products and they should exchange it. I walked into a tan and orange box store, they said no problem, go find one on the shelf and we'll just exchange it. They did. May not be the top of the line, but that's great customer service. Now if Ford and Chevy would do the same thing.....
SlingBob is offline   Reply With Quote
Old 01-12-2017, 11:46 AM   #14
Deplorable-Electrician
 
MechanicalDVR's Avatar
 
Join Date: Dec 2007
Location: God's Country of Virginia
Posts: 31,237
Rewards Points: 2,108
Default

Quote:
Originally Posted by SlingBob View Post
Klein suggested taking it to any vendor that deals in the products and they should exchange it. I walked into a tan and orange box store, they said no problem, go find one on the shelf and we'll just exchange it. They did. May not be the top of the line, but that's great customer service. Now if Ford and Chevy would do the same thing.....
That reminds me of Snap-on sockets being guaranteed for life. I have broken plenty of them and had them replaced free of charge. Then there was Craftsman, also lifetime warranty and about 1/10th the price and I've only broken like one.

Top of the line is often skewed. You lucked out in the way they handled it.

I'd never take the chance on a Klein meter, nor would I buy a Fluke screwdriver.
__________________
I have never claimed to be a nice guy!

"Injustice anywhere is a threat to justice everywhere" mlk
MechanicalDVR is offline   Reply With Quote
Reply

Thread Tools
Display Modes

Posting Rules
You may not post new threads
You may not post replies
You may not post attachments
You may not edit your posts

BB code is On
Smilies are On
[IMG] code is On
HTML code is Off
Trackbacks are Off
Pingbacks are Off
Refbacks are Off


Similar Threads
Thread Thread Starter Forum Replies Last Post
Are You Lost? Ask An ElectricianTalk Moderator Cricket Site Help and Suggestions 223 05-01-2017 05:38 PM
GIVEAWAY: Klein Tools 25 Ft. Double Hook Magnetic Tape Measure Cricket General Electrical Discussion 77 12-16-2016 07:24 PM
PSA: Milwaukee Stand Light clamp fix UncleMike Tools, Equipment and New Products 0 08-19-2016 08:27 PM
Union j-man's tool list MechanicalDVR Tools, Equipment and New Products 57 05-27-2016 10:59 AM
Non contact and clamp meter suggestions? 120ToTheFingers Tools, Equipment and New Products 21 05-25-2016 09:50 PM


All times are GMT -4. The time now is 04:45 PM.


Powered by vBulletin® Version 3.8.8
Copyright ©2000 - 2017, vBulletin Solutions, Inc.
Search Engine Optimization by vBSEO 3.6.1
vBulletin Security provided by vBSecurity v2.2.2 (Pro) - vBulletin Mods & Addons Copyright © 2017 DragonByte Technologies Ltd.
User Alert System provided by Advanced User Tagging (Pro) - vBulletin Mods & Addons Copyright © 2017 DragonByte Technologies Ltd.
Our Pro Sites Network
ContractorTalk.com | DrywallTalk.com | HVACSite.com | PaintTalk.com | PlumbingZone.com | RoofingTalk.com